Company record
Auris Health, Inc
0 FDA warning letters since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about Auris Health, Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Feb 25 | Class II | Auris Health, Inc Device |
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube. FDA notice Z-1383-2026 |
| Nov 26, 2025 | Class II | Auris Health, Inc Device |
MONARCH Bronchoscope. Model Number: MBR-000211-B Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks. FDA notice Z-0543-2026 |